Study Outline
Porous packages are prepared by placing tape over the porous surface to reduce the rate which air escapes. The set up packages are used to set the machine to 80% of the lowest burst pressure as determined from a previous burst test. A probe is inserted into the sample package and the machine started. The package inflates to the set pressure and the timer starts. The test continues until the package bursts (failure) or the set time is reached. The machine records the pressure for each sample. Any failures are reported with time to failure and failure location.
Sample Requirements
Minimum sample size: complete packages
Recommended replicates: The number of test packages that are needed for this test are at the discretion of the sponsor. However, Nelson Labs recommends that at least 11 test packages be tested to make sure that the test is statistically significant.
Minimum per product type/lot: 11 samples + 3-5 set-up samples